Opened cassette with slotted inlet facing up
Filter cut in half between two sample traces
|
A "key and slot" is included in the perimeter
of the cassette, so that the cassette only fits together one way. This
allows field personnel to remove one trace for microscopic analysis, re-assemble
the cassette, and send the second trace to a laboratory for culturing
or other analysis. If the small end cap is removed, the large end cap
may be used to support the cassette upright.
- Cut through the label securing the two halves of the cassette
- Position the cassette so that the slotted inlet is facing up
- Remove the top portion of the cassette
- Place the filter, on the support pad, onto a clean surface
- Cut the filter in half between the two sample traces
|
|
1/2 Filter on slide before clearing
1/2 Filter on slide after clearing
|
Microscopic Analysis: MCE filter + Acetone vapor
- Place one sample trace (half-filter) on a glass slide
- Insert the half-filter and slide into an acetone vaporizer and clear
the filter
- Place a drop of 3% KOH or similar liquid on the cleared filter
- Cover the sample with a cover slip
- Analyze the sample trace at a minimum magnification of 600X using
light microscopy
|